Decontamination works when soil is removed completely: clean every surface, verify the result, and follow the manufacturer IFU at each step. These 15 free practice questions drill the decontamination topics behind 21% of the CRCST exam, and every answer cites its source.
Updated October 2026
Work through all 15 questions below. Every answer includes an explanation that cites its source. When you finish, head back to the full 50-question practice test.
Decontamination
Question 1: The decontamination area is kept cooler than the rest of the department, around 60 to 65 degrees F (16 to 18 degrees C). What is the main reason?
Correct answer: B
Warm, humid rooms accelerate bacterial growth on soiled instruments waiting for processing, and decontamination PPE (fluid-resistant gown, heavy gloves, face shield) is exhausting in heat. AAMI ST79 recommends the cooler range for the decontamination area. (Source: AAMI ST79)
Decontamination
Question 2: For manual cleaning with an enzymatic detergent, what water temperature should the sink be?
Correct answer: B
Water that is too hot denatures enzymes and bakes protein soil onto instruments, while cold water reduces cleaning action. The detergent IFU gives the temperature window where its enzymes work best, typically warm or tepid. (Source: AAMI ST79)
Decontamination
Question 3: Before an ultrasonic cleaner is used for the first patient load of the day, the technician must degas it. How is degassing done?
Correct answer: B
Fresh solution holds dissolved air that absorbs ultrasonic energy and weakens cavitation. Running the unit empty for the IFU stated time drives the trapped air out so the first load gets full cleaning power. (Source: AAMI ST79)
Decontamination
Question 4: When loading instruments into an ultrasonic cleaner, which practice is correct?
Correct answer: B
Cavitation must reach every surface. Instruments touching each other or the tank create shadow zones where bubbles cannot form, and items above the solution get no cleaning at all. Baskets keep the load separated and fully immersed. (Source: AAMI ST79)
Decontamination
Question 5: A washer-disinfector achieves thermal disinfection with which minimum time and temperature combination?
Correct answer: C
Thermal disinfection in a washer-disinfector is defined by time-temperature relationships, and 194 degrees F (90 degrees C) for 1 minute is the widely recognized minimum. Higher temperatures allow shorter times per the equipment IFU. (Source: AAMI ST79)
Decontamination
Question 6: After manual cleaning, how can a technician verify that instruments are actually clean before they move to assembly?
Correct answer: B
Clean to the eye is not clean to the microbe. Residue tests detect invisible protein, blood, or organic soil left on surfaces and inside lumens, proving the cleaning process worked before instruments advance to packaging. (Source: AAMI ST79)
Decontamination
Question 7: Small pits are forming on stainless steel instruments. What is the most common cause?
Correct answer: B
Chloride ions attack the chromium-oxide passive layer that protects stainless steel, and dried saline concentrates the attack in one spot. Instruments exposed to saline should be rinsed promptly so chlorides never sit on the surface. (Source: AAMI ST79)
Decontamination
Question 8: Aluminum instrument containers come back from the washer with dark discoloration. What likely caused it?
Correct answer: A
Aluminum is attacked by highly alkaline detergents, which etch and darken the metal. Aluminum items need neutral detergents and processing per the container manufacturer IFU. (Source: AAMI ST79)
Decontamination
Question 9: Why must instruments be thoroughly rinsed after cleaning and before disinfection or sterilization?
Correct answer: B
Detergent films bake onto surfaces in the sterilizer, causing staining and creating a barrier between the sterilant and any remaining soil. A complete rinse, finished with treated water, leaves nothing between the instrument and the sterilant. (Source: AAMI ST79)
Decontamination
Question 10: Reusable cleaning brushes must be:
Correct answer: B
A worn or contaminated brush redeposits soil instead of removing it, and damaged bristles can scratch instruments or break off inside lumens. Brushes are cleaned after each use, inspected, and retired when bristles splay, loosen, or show damage. (Source: AAMI ST79)
Decontamination
Question 11: After manual cleaning of a cannulated (lumened) instrument, the technician must:
Correct answer: B
Lumens hide soil where eyes cannot follow. Flushing with a syringe and watching for free, unobstructed flow confirms the channel is open and clean; a blocked or sluggish lumen goes back for re-cleaning. (Source: AAMI ST79)
Decontamination
Question 12: A technician figures that doubling the detergent will clean twice as fast. What is wrong with this idea?
Correct answer: B
Detergents are formulated to work at a specific concentration. Overdosing leaves films that resist rinsing, and the residue can stain instruments and block sterilant contact, the opposite of the intended result. (Source: AAMI ST79)
Decontamination
Question 13: AAMI ST79 recommends a minimum of how many air exchanges per hour in the decontamination area?
Correct answer: C
Decontamination generates aerosols, humidity, and chemical vapors from detergents and disinfectants. Ten air exchanges per hour dilute contaminants and carry moisture out, protecting both staff and the cleaning process. (Source: AAMI ST79)
Decontamination
Question 14: After finishing manual cleaning, in what order should the technician remove PPE?
Correct answer: B
Doffing starts with the most contaminated item, the gloves, and ends with the mask, which protected the airway throughout. Removing PPE before leaving the decontamination area keeps contaminants from traveling into clean zones. (Source: OSHA 29 CFR 1910.1030)
Decontamination
Question 15: Gross soil should be removed from instruments before they go into the ultrasonic cleaner because:
Correct answer: A
Ultrasonic cleaning is a fine-cleaning step, not a substitute for gross soil removal. Thick blood and tissue dampen the cavitation bubbles, so pre-rinsed instruments come out far cleaner than ones loaded straight from the OR. (Source: AAMI ST79)
Every explanation on this page cites one of the references below. Always follow your facility's policies and the manufacturer instructions for use.