Steam sterilization works when saturated steam touches every surface for the full exposure time at the right temperature. These 15 free practice questions drill sterilization parameters, monitoring, and low-temperature methods, with every answer cited.
Updated October 2026
Work through all 15 questions below. Every answer includes an explanation that cites its source. When you finish, head back to the full 50-question practice test.
Sterilization
Question 1: In a pre-vacuum (dynamic air removal) steam sterilizer at 275 degrees F (135 degrees C), the minimum exposure time for wrapped instrument trays is:
Correct answer: B
Higher temperature means shorter exposure. At 275 degrees F (135 degrees C), AAMI ST79 lists 3 minutes as the minimum exposure for wrapped goods in a pre-vacuum cycle, with drying time afterward. (Source: AAMI ST79)
Sterilization
Question 2: For gravity-displacement steam sterilization of wrapped instruments at 270 degrees F (132 degrees C), the minimum exposure time is:
Correct answer: B
Gravity cycles remove air passively, so they need longer exposures than pre-vacuum cycles at the same temperature. At 270 degrees F (132 degrees C), wrapped goods require 15 minutes of exposure. (Source: AAMI ST79)
Sterilization
Question 3: For dry-heat sterilization of wrapped instruments at 340 degrees F (171 degrees C), the minimum exposure time is:
Correct answer: C
Dry heat kills by oxidation, which is slower than moist heat, so exposures run much longer. At 340 degrees F (171 degrees C), wrapped instruments need 1 hour of exposure. (Source: AAMI ST79)
Sterilization
Question 4: Biological indicators used to monitor ethylene oxide sterilization contain spores of:
Correct answer: B
Each sterilization method gets a BI organism chosen for its resistance to that method. Bacillus atrophaeus spores resist ethylene oxide, while Geobacillus stearothermophilus is the steam and hydrogen peroxide indicator organism. (Source: AAMI ST79)
Sterilization
Question 5: Which item must NEVER be processed in a hydrogen peroxide gas plasma sterilizer?
Correct answer: B
Hydrogen peroxide gas plasma cannot penetrate cellulose-based materials, which absorb the sterilant, and liquids or powders are also incompatible. Only devices and packaging validated for the specific system may be processed. (Source: AAMI ST79)
Sterilization
Question 6: The Bowie-Dick test is used only on dynamic air removal (pre-vacuum) sterilizers, not gravity sterilizers, because:
Correct answer: B
The Bowie-Dick test pack is an air-removal challenge. Gravity sterilizers do not mechanically remove air, so the test has nothing to challenge; it belongs to pre-vacuum cycles where trapped air is the failure mode it was designed to catch. (Source: AAMI ST79)
Sterilization
Question 7: When loading a steam sterilizer, packs and trays must be positioned so that:
Correct answer: B
Sterilization happens where steam touches. Packs pressed against walls or each other create cold spots where air and condensate collect, so loads are arranged with space for circulation and kept off the chamber walls. (Source: AAMI ST79)
Sterilization
Question 8: After the exposure phase of a steam cycle ends, wrapped trays need:
Correct answer: B
Exposure kills microorganisms, but the cycle is not finished until the load is dry. Opening the door early or skipping dry time leaves damp packages, and a wet pack is a contaminated pack. (Source: AAMI ST79)
Sterilization
Question 9: Superheated steam is undesirable in a steam sterilizer because:
Correct answer: B
Sterilization by steam depends on moisture carrying heat into proteins and denaturing them. Superheated steam behaves more like dry hot air, which kills far more slowly, so sterilizers are designed to deliver saturated steam at 97 to 100 percent quality. (Source: AAMI ST79)
Sterilization
Question 10: The steam trap (thermostatic trap) on a steam sterilizer exists to:
Correct answer: B
Condensate pooling in the chamber creates wet packs, and trapped air blocks steam contact. The trap continuously drains water and vents air so only dry, saturated steam reaches the load. (Source: AAMI ST79)
Sterilization
Question 11: Physical monitoring of every steam sterilization cycle requires verifying and documenting:
Correct answer: B
Time, temperature, and pressure are the physical critical variables of steam sterilization. The sterilizer printout or digital record must show all three stayed within specifications for the full exposure. (Source: AAMI ST79)
Sterilization
Question 12: A biological indicator comes back positive (growth). What must happen next?
Correct answer: C
A positive BI means sterilization conditions were not achieved. The sterilizer is taken out of service until the cause is found and corrected, and every load back to the last negative BI is recalled, because none of those loads can be trusted. (Source: AAMI ST79)
Sterilization
Question 13: Instruments processed by immediate-use steam sterilization (IUSS) must be:
Correct answer: B
IUSS skips the packaging and storage safeguards of terminal sterilization, so its products have no sterile barrier. The items go directly to the point of use; anything not used is reprocessed, never shelved. (Source: AAMI ST79)
Sterilization
Question 14: A Type 6 chemical indicator is best described as:
Correct answer: B
Type 6 emulating indicators are cycle-specific: they react to every critical variable of one defined cycle, such as 4 minutes at 270 degrees F in pre-vacuum steam. They are the most cycle-precise chemical monitors available. (Source: AAMI ST79)
Sterilization
Question 15: A process challenge device (PCD) used with a biological indicator is designed to:
Correct answer: B
A PCD places the BI inside a test pack that mimics the most resistant spot in the load, such as the center of a dense tray. If sterilant reaches and kills the BI there, the rest of the load can be trusted. (Source: AAMI ST79)
Every explanation on this page cites one of the references below. Always follow your facility's policies and the manufacturer instructions for use.
Not affiliated with or endorsed by HSPA.